Technical specifications
Modes
CPAP,
S, T, ST with optional iBR, PAC, iVAPS with iBR and optional AutoEPAP
Operating pressure range
2–25
cm H2O (2–25 hPa) in S, T, ST, PAC and iVAPS modes
4–20
cm H2O (4–20 hPa) in CPAP mode
Sound pressure level
26.6
dBA with uncertainty of 2 dBA as measured according to ISO 17510-1:2009 (CPAP
mode)
Dimensions (H x W x D)
116
mm x 205 mm x 150 mm (Device only)
116
mm x 255 mm x 150 mm (Device with HumidAir humidifier)
Weight
1106
g (Device only)
1268
g (Device with HumidAir humidifier)
90W power supply unit
Input
range: 100–240V, 50–60Hz, 57VA (typical power consumption) 108VA (peak power consumption)
Operating temperature/humidity/altitude
+5°C
to +35°C/ 10–95% relative humidity, non-condensing/ Sea level to 2,591 m; air
pressure range 1013 hPa to 738 hPa
Storage and transport temperature/humidity
-20°C
to +60°C/ 5–95% relative humidity, non-condensing
Housing construction
Flame
retardant engineering thermoplastic
Supplemental oxygen
Recommended
maximum flow: 15 L/min
(CPAP,
S, T, ST, PAC); 4 L/min (iVAPS)
Standard air filter
Polyester
non-woven fibre
Tubing
SlimLine™
and ClimateLineAir (15 mm); Standard (19 mm)
Air outlet
22
mm air outlet complies with ISO 5356-1:2004
Electromagnetic compatibility
Requirements
(EMC) according to IEC60601-1-2:2007 for residential, commercial, and light
industry environments
Aircraft use
ResMed
confirms that device meets the Federal Aviation Administration (FAA)
requirements
(RTCA/DO-160,
section 21, category M) for all phases of air travel
IEC 60601-1: 2006 classification
Class
II (double insulation), Type BF Ingress protection IP22