Breas Vivo 45 LS
Breas Vivo 45 LS
- Brands Breas
- Product Code: Breas Vivo 45 LS
- Availability: Out Of Stock
- Rs.554,400
- (including GST & Delivery charges)
Vivo 45 LS Technical Specifications
Settings / Performance | |
---|---|
Ventilation Modes | PSV, PSV(TgV), PCV, PCV(TgV), PCV(A), PCV(A+TgV), PCV-SIMV, PCV-MPV, VCV, VCV(A), VCV-SIMV, VCVMPV, CPAP |
Patient Modes | Adult, Pediatric |
Device Modes | Home, Clinical |
Profiles | 3 settable profiles |
Inspiratory Pressure | 4 to 50 cmH2O, max PS of 30 cmH2O above PEEP (Pediatric) |
PEEP | 2 cmH2O to 20 cmH2O, IPAP -2 cmH2O or Min IPAP -2 cmH2O |
Breath Rate/Backup Rate | 4 - 40 bpm (Adult), 6 - 60 bpm (Pediatric) |
Inspiratory Time/Backup Insp. Time | 0.3 to 5 s (Adult), 0.3 to 2 s (Pediatric) |
Rise Time | 1 to 9 (PSV & PCV), 50 to 90%, Off (VCV) |
Inspiratory Trigger | 1 to 9 (PSV & PCV, VCV), Off (PCV, VCV) |
Expiratory Trigger | 1 to 9 |
Minimum Inspiratory Time | Off, 0.3 to 3 s (Adult), Off, 0.3 to 2 s (Pediatric) |
Maximum Inspiratory Time | 0.3 to 5 s, Off (Adult), 0.3 to 2 s, Off (Pediatric) |
Tidal/Target Volume | Off, 300 - 2000 ml (Adult), Off, 50 - 400 ml (Pediatric) |
Monitoring | |
Displayed data | Ppeak, PEEP, Pmean, Leakage, MV, Vt, FiO2, % in TgV, Total Rate, Spont Rate, % Spont, SpO2, Pulse Rate, I:E, Inspiratory Time, Rise Time, etCO2, InspCO2, PtcCO2 |
Waveforms | Pressure, flow, volume, etCO2, leakage |
Trends over 1, 6, 24, 48 h and 10 days on device screen | SpO2, Ppeak, PEEP, Total Rate, Spontaneous Rate, Leakage, Vte, etCO2 |
Power Supplies | |
Mains supply | 100 to 240 V AC |
External DC | 19 V ± 6 V |
Click-in battery | 5.5 hours |
Internal battery | 2 hours |
Main alarms | |
Alarms | High Pressure, Low Pressure, High PEEP; Low PEEP, High Vti, Low Vti , High MVi, Low MVi, High Vte, Low Vte, High MVe, Low MVe, High Breath Rate, Low Breath Rate, Apnea, Disconnection, Rebreathing, Obstruction, High FiO2, Low FiO2, High SpO2, Low SpO2, High etCO2, Low etCO2, High InspCO2, High Pulse Rate, Low Pulse Rate, PtcCO2, Power Fail, High Patient Air Temp, Low Patient Air Temp, Low Last Power Source, Crit. Low Last Power Source, Lost Mains Power, Exhalation Valve Control Error, SpO2 Disconnection, SpO2 Signal Lost, Poor SpO2 Signal, CO2 Disconnection,CO2 Accuracy, Check CO2 Adapter, CO2 Sensor, FiO2 Disconnection. |
Dimensions | |
W x H x D | 8.50" x 6.26" x 5.98" (216 x 159 x 152 mm) |
Weight | 5.3 lb (2.4 kg) |
Noise level (at 10 cmH2O constant pressure) | Less than 30 dB(A) |
Connectivity and Software | |
Built-in connectivity | Ethernet, USB |
Compatible Software | EveryWare by Breas |
Plug-and-Play accessories | SpO2, PtcCO2, etCO2, FiO2, Nurse Call, Remote Alarm, Remote Start/Stop |
Other | |
Circuits | Leakage Circuit, MPV Circuit, Circuit with Active Exhalation Valve |